EL041
EnviroMon Brochure
EnviroMon User Guide
Get your EnviroMon quotation today.
Related Links
EnviroMon Converters & Sensors |
EnviroMon
(All trademarks acknowledged) E. & O. E.
|
The 4 channel EL041 thermocouple converter is designed to measure a wide range of temperatures with any thermocouple that uses a miniature size thermocouple connector.
Featuring built-in cold junction compensation (CJC), the EL041 has an effective temperature range of -270 to 1820°C (The actual temperature range depends on the thermocouple being used.)
In addition to temperature measurement, the cold junction compensation can be switched off to enable the EL041 to measure ±60mV signals.
Note: The EL041 is designed to work only with the EL005 logger.
EL041 Thermocouple Converter Specifications |
Number of channels |
4 |
Thermocouples Supported |
B, E, J, K, N, R, S, T or mV |
Voltage Input Range |
±60 mV |
Resolution |
>15 bits + sign |
Accuracy |
Thermocouples: ±(0.5 °C + 0.3% of type K reading) Millivolts: ±0.3% |
Conversion Time |
1 second |
Overload Protection |
±10 V |
Input Connector |
Mini-thermocouple plugs |
PC Connector |
via EnviroMon network |
Operating Temperature |
0 to 70 °C (20 to 30 °C for quoted accuracy) |
Operating Humidity |
25 to 75%RH |
Ask for EL041 detailed specification.
|